Velocity reference standards for predicting stroke in children with sickle cell disease were obtained with non-imaging TCD equipment. Distinct equipment is used for TCD and TCDI, but both analyze spectral Doppler waveforms to assess flow.
- Transcranial Doppler Ultrasound Equipment (TCD)
- Utilizes a single crystal 1.5-2.5MHz pulsed Doppler probe
- Equipment and software specifically designed for TCD applications
- Displays all data in real time including:
- Sample volume, sample depth
- Time averaged maximum mean velocity
- Peak systolic velocity/end diastolic velocity
- Acceleration time
- Pulsatility (PI) or resistive indices (RI)
- Doppler power output and frequency
- Direction sensitive Doppler blood flow meter
- Bi-directional Doppler waveform display
- Audible output and permanent recording of the spectral waveform
- Transcranial Color Duplex Imaging Equipment (TCDI)
- Duplex imaging scanner with B-Mode, color and spectral Doppler capabilities
- Utilizes a 1-5MHz transducer probe
- Displays all data in real time including:
- Sample volume, sample depth, Doppler angle
- Time averaged maximum mean velocity
- Peak systolic velocity /end diastolic velocity
- Acceleration time
- Pulsatility (PI) or resistive indices (RI)
- Doppler power output and frequency
- Direction sensitive Doppler blood flow meter
- Bi-directional Doppler waveform display
- Visual display, audible output and permanent recording of the spectral waveforms with corresponding B-Mode/Color image